UbiqVue™ Holter Analyzer Server and Desktop

UbiqVue™ Holter Analyzer Server and Desktop



Internal Release: 4.1.0.173

Bug fixes

Internal Release: 4.1.0.161

Heart Rate & Beat Distribution Updates

  • Updated Heart Rate and Beat Distribution tables on the report’s front page.
  • Tachycardia, Bradycardia, and Pause thresholds are now pulled from user settings and reflected in the beat distribution table.
  • Triplet counts are now included in the SVT table.
  • SVT total counts corrected in the beat distribution table.
  • Burden percentages for SVC/PACs and VE/PVCs now use only classified beats, improving clinical accuracy.

Patient & Rhythm Event Handling

  • Resolved issues related to ECG technician data entry and processing.
  • Enhanced support for version 7 patient XML files, including:
  • Parsing updated patient event data format (e.g., button presses).
  • Reading and applying ST segment thresholds.
  • Removed Patient Diary attribute from the Patient Event Table.
  • Corrected Tachycardia and Bradycardia beat counts, now based on user-defined thresholds.

Graphical & Rhythm Strip Improvements

  • 6-second graphs now display Heart Rate in an avg [min–max] format.
  • Longest SVT duration is now shown on the VT 6-second graph.
  • AFib/Flutter selection logic improved:
  • A Flutter strip will be shown if AFib is not detected.
  • Tachycardia, Bradycardia, and Normal Sinus rhythm strips are now included in the Additional section of the report.
  • AFib burden now displays the % symbol as per the updated report template.

Visual & Layout Enhancements

  • Fixed signature overlap in the Interpreter Posted Final Report.
  • Adjusted alignment and text wrapping for improved report readability.

Internal Release: 4.1.0.130

Initial Release



Version H

Release Date: 13-Aug-2024

System Upgrades:

  • Security updates

System Upgrades:

  • Integration of connected devices for additional parameters: ContecMed Wrist/Finger SpO2, Omron Blood Pressure Monitor
  • Added ECG trends for more comprehensive monitoring
  • 976 samples per second support for advanced biosensors
  • Support for UbiqVue 1AX Biosensor and associated parameter monitoring
  • Added compatibility features to support NALAM Hospital Management Software

User Experience Improvements:

  • Added PHI consent prompt on first-time login
  • Improved UI and removed unnecessary checkboxes to streamline alert configuration
  • Enhanced audit trail functionality to log all user actions during logged-in sessions
  • Threshold adjustments for temperature and SpO2 alerts.

Bug Fixes:

  • Resolved an issue with alert times not updating on the dashboard.
  • Fixed skin temperature alerts not deactivating when values returned to normal
  • Addressed SpO2 alert forwarding for specific conditions
  • Removed hardcoded S3 bucket for improved system flexibility

Initial Release:

  • Support for UbiqVue 2A Biosensor and associated parameter monitoring