UbiqVue™ Holter System Anomalies and Workarounds

UbiqVue™ Holter System Anomalies and Workarounds

Updated:

ComponentAnomalyWorkaround
ComponentUbiqVue™ Holter Web PortalUA2251-C Anomaly
  • For Data Retrieval Clinician users, the EDF may display the logged-in user’s time zone instead of the patient’s time zone.
Workaround—
ComponentUbiqVue™ Holter Analyzer Server and DesktopUA2251-SA, UA2251-DA Anomaly
  • Report upload may fail in certain multi-patch workflows when the ECG Technician and Final Interpreter are different.
WorkaroundLog out and log back in, then retry the report upload.
ComponentUbiqVue™ Holter Analyzer Server and DesktopUA2251-SA, UA2251-DA Anomaly
  • A software crash may occur in the Insert New Beat workflow if selected beats are removed before the initial search for matching beats is completed.
WorkaroundRun Find first and allow the search to complete. Then return to the selected beats list and right-click any beats you want to remove. Add additional beats as needed and run Find again.